61tH dAy Of AtTaChMenT
There were much repair to do for the first two weeks of repair and preventive maintenance can be done in 2 weeks. Repair for intensive care unit and emergency ward seems like the first priority. Two intermittent suction units were sent down to the workshop. Both of them having the same problem which is no suction for intermittent mode. While i analyzed the problem, i found out that there is suction for continuous mode. I look through the service manual and found out the timing and flow control valves were turn to the extreme clockwise from the troubleshooting section. I attempted to adjust the valves but the problem still exist.
I got the chance to see the older model of defibrillator and vital sign spot monitor. There is a fault for the vital sign spot monitor which is the SpO2 readings could not be shown. Initially, it is unable to record the reading. After a few attempts, i find that the reading will not be immediated and takes longer time compared to the later model. I took the service manual as i am not sure how to go to the service mode to adjust the time as the time is off. I got to more things about servicing compared to the past.
I got the chance to see the older model of defibrillator and vital sign spot monitor. There is a fault for the vital sign spot monitor which is the SpO2 readings could not be shown. Initially, it is unable to record the reading. After a few attempts, i find that the reading will not be immediated and takes longer time compared to the later model. I took the service manual as i am not sure how to go to the service mode to adjust the time as the time is off. I got to more things about servicing compared to the past.