5 months of attachment

Wednesday, May 31, 2006

55Th DaY of AttachMent

There were a couple of repair in the morning. There were a nebuliser and sphygonomometer being sent down. Both of them seems rather simple repairs as they involved in changing the certain parts of the devices.

Between 10 am to 11 am, i followed Faz to emergency medicine to mount brackets for 11 propaq as sister wanted the propaq to be able to place on a metal pole for 2 reason which are taking less place and easy to transport from a place to another. However, it is time consuming to do all 11 propaq at one go.

Around 2:30 pm, Elson from Mandarin came to do commissioning for 4 equipments which 2 of them are for trial and the others are being bought by us. One of the is the latest model slit lamp, BQ 900, and the other is laser unit, IRIDEX. For the slit lamp, they just do the physical checks like the brightness and intensity of the light being shined onto the eye. He did the power measurement test for the laser unit. The tolerance for the laser unit is about +/-20 % of the stated values ( 200W, 500W, 1000W, 2000W). Both of them would have safety test required. Slit lamp is for diagnosing of human eye while the laser unit is used to perform ophthalmic surgery.

0 Comments: