We went to Clinic C and J to collect the electrical sphygonomamometer (NIBP) for preventive maintenance as there are quite a number of them due this month. Besides the safety leakage current test, pressure, temperature and accessories tests are to be done necessarily inclusive of battery test. A pressure tester is required to check for the accuracy of pressure of the device. Temperature test is carried out by putting the temperature probe into a cup of warm water and using thermometer to measure the temperature.
Around 10 am, zhong hong's colleagues and him came by to my workplace to deliver sleep studies equipments as they are bought for research. However, not all the equipments are received so we were unable to do the commissioning for it. It was nice to see him again after 3 months since i last saw him.
In the afternoon, we went to Clinic C to do a thorough clean-up of indirect opthalmoscopes. There are about 6 indirect opthalmoscopes in Clinic C. We wiped the lens with tissue and small pump to blow the dust away.
Around 10 am, zhong hong's colleagues and him came by to my workplace to deliver sleep studies equipments as they are bought for research. However, not all the equipments are received so we were unable to do the commissioning for it. It was nice to see him again after 3 months since i last saw him.
In the afternoon, we went to Clinic C to do a thorough clean-up of indirect opthalmoscopes. There are about 6 indirect opthalmoscopes in Clinic C. We wiped the lens with tissue and small pump to blow the dust away.