5 months of attachment

Friday, August 11, 2006

106th day of attachment

Today is my last day of work and i really enjoyed myself during my experience in Alexandra Hospital. I finished my report and let my supervisor to look through it. I really like to thanks every Biomedical engineering staffs to make my experience so wonderful. I really learnt quite a lot of things from them.

Thursday, August 10, 2006

105th day of attachment

I continue doing my report and help up the administrative staff to do paper work.

Tuesday, August 08, 2006

104th day of attachment

In the morning, i continue doing checking for the other 2 devices that are requested as loan. Initially, I just connect the resuscitator with the oxygen cylinder and the patient valve to check for its working condition. Later, the flow analyzer was used to check the accuracy of the tidal volume that it produced and breath per minute. My colleague and I just simply check the defibrillator by having physical check. We making sure of the paddles' working condition. For Lifepak 12, it is more sophisticated compared to Lifepak 20 as it has more functions. At 3am, it will do a self – test automatically so pacer not recommended to be connected to the device unless it required to be used. After the staff who is in charge of this event came to collect the devices, i continue doing my report and a bit of my presentation.

In the afternoon, alternating mattress pressure system, AlphaxCell is sent down for repair. The fault lies with the parts of the connector stuck inside of the connector of the mattress. We solved the problem by removing the part of the connector instead of replacing with a new one. An suction regulator is brought back for in-house maintenance. Vacutron intermittent/continuous suction regulator from Allied Healthcare. This suction regulator is different from all of the suction regulators that I have done either repair or maintenance for. However, the functions still the same. The intermittent timings for this model having shorter range of time. The staff who is in charge for helping out in the marathon requested for an oxygen transport flowmeter connected with another small oxygen cylinder. Before loaning it to him, I have to check whether it is in working condition. Since there is no service manual or functional check list given to us upon purchase, we only able to roughly gauge whether it is still working.

Monday, August 07, 2006

103th day of attachment

In the morning, I did preventive maintenance for the Thoraric Drainage aspirator and ear thermometer as they are due since last month. After that, i continue editting my final report before letting my supervisor to read through it.

In the afternoon, I done a bit of repair since my colleague is on leave. I repaired a intermittent suction regulator and analyze the problem with the alternating mattress pump cotrol from betabed. After that, i was being assigned a task which involved in checking of devices required for the marathon event, Philips Singapore Cup (11-13 August) . However, I only managed to complete the checking and preventive maintenance for transport physiologic monitor, Propaq 202EL.