5 months of attachment

Thursday, May 04, 2006

37th DaY oF aTtAcHmEnT

Searching from high to low for the infusion pumps from ward 1, 2 and 10. Thankfully, i managed to find the infusion pump from ward 1 and 2 but for ward 10, i found a infusion pump that was due last month so brought back to do preventive maintenance. MICU sent down an infusion pump for preventive maintenance.

I managed to finish the battery, occlusion and flow test before lunch. However, the leakage current test only able to be done after lunch as Faz brought the safety analyzer with her for PM. while i was doing a flow test for STC 508, the whole department suddenly short circuit for a moment and i have to redo the flow test as the infusion device analyzer does not have internal battery.

There was a commissioning for tonometer which is an equipment on trial for a few months. The vendor will assemble the device and test the functions according to the checklist. The vendor will use their eyes for the function test. There is a tonometer which looks like a pen (tonopen) has the same property as this device. The only difference is tonopen has to be placed directly on the cornea whereas the tonometer has a distance apart from the camera and jet of air will be delivered to the eye when recording the measurement.