5 months of attachment

Tuesday, May 02, 2006

35th DaY of AtTaChMent

There were pretty many devices that sent for repair including the laderal suction unit also known as portable aspirator. Today is also the first day of may so a new month of preventive maintenance starts again. pretty excited as there seems many more devices for me to handle. I helped out by writing the class type for each individual devices for the leakage test as usual. However, i kind of doing administrative work more than doing the preventive maintenance as Norainie is not around so i just opened out work orders and answering and making calls.

Got back the sphygnomamometer back and this time, the nurse assumed that the mercury dried up but the fact is they did not unlock the bottle in order for mercury to flow. However, in the afternoo, i did some inspection of devices that were known as faulty and their accessories. saw the nihon korden's physiologic monitor as the problem is that the cuff is unable to be pump until 120 mmHg at least. It is due to the the setting of the pressure limit is too low. After adjusting the setting, the pressure is able to reach until 170 mmHg. Lee Chong suspected one of the end user tends to adjust the setting pretty often.

Tomorrow will be pump day for me as i will be doing preventive maintenance for infusion and syringe pump while Faz is not around for the afternoon and have training for he blood analyzer from IDS in the morning.